Photon counting histogram: one-photon excitation

Chemphyschem. 2004 Oct 18;5(10):1523-31. doi: 10.1002/cphc.200400176.

Abstract

The photon counting histogram (PCH) analysis is a fluorescence fluctuation method that is able to characterize the brightness and concentration of different fluorescent species present in a liquid sample. We find that the PCH model using a three-dimensional Gaussian observation volume profile is inadequate for fitting experimental data obtained from a confocal setup with one-photon excitation. We propose an imoroved model, which is based on the correction to the observation volume profile for the out-of-focus emission. We demonstrate that this model is able to resolve different species present under a wide range of conditions. Attention is given to how this model allows the examination of the effects of different instrumental setups on the resolvability.